Wednesday 27 January 2016



Acoustic lens damages in ultrasound probes



Probes are the soul part of all ultrasound imaging systems, because the performance and imaging quality of ultrasonic scanner are highly affected by the characteristic and the structure of the probe. This hand held device often gets damaged due to its constant use. Thus, proper care and regular testing of probes significantly influence the level of operating expenses of the ultrasound systems. 

Common acoustic lens failures

        1.  Holes in lens


    
This is one of the common lens damage.
    The reasons for this damage are:
  •          Improper cleaning
  •          Rough handling, dropping probes from hand
  •          Use of gels having perfumes or other molecular changing substances.

  • lens with multiple holes

    lens with single hole














           
\       Holes in lens can lead to delamination
       
        2. swollen lens
        
          

















    This is also happening due to the improper cleaning and repeated use. This may leads to delamination and will not get ultrasound image.
   
     3.    Lens separated from case


    












   
     This is occurring due to Improper or insufficient cleaning, or cleaning with wrong substance i.e, alcohol-based wipes on the lens tends to dry the lens and make it detach from the probe.

4    4.    Lens missing


     



    










    
     The only solution for acoustic lens damages is lens replacement.
     For more  information on this visit www.probelab.in or call 9847069684





    







   

Wednesday 6 January 2016



NECESSITY OF VALIDATION OF PROBES BEFORE 

WARRANTY EXPIRES

The main problem associated with all ultrasound probes is that they have a limited life span or warranty period. Due to their repeated and extended use over a time period, they get damaged easily. An out-of-date ultra sound probe may fail to function to manufacturer's specifications, and can cause health risk to the patient.
There are many reasons for probe damage. Some of them are:       
                    
  •       Drop from hand : leads to crystal damage
  •       Improper sterilization and disinfection          process : leads to de-lamination and oil leak
  •       Rough handling : causes cable cut, connector damage

The proper functioning of the ultrasound transducer is a key factor for reliable diagnosis by ultrasound. This is highly depends upon the working of the piezoelectric crystals, the wires within the transducer ,matching layer in front of the elements and the backing material behind the elements. Hence they are the internal parts; it is very difficult to identify any damages to the system until the picture quality gets affected. In most cases such damages to the transducers leads to:

·        False diagnosis and treatments
·        Loss of revenue
·        Loss of reputation in market
·        Patient safety problems  etc
     Now days by using advance probe testing technique like first call test, we can test all essential transducer parameters according to FDA regulation. And can easily identify any damages to the elements in the ultrasound transducers. (First call test of an E8C RS probe is available at this think https://www.youtube.com/watch?v=DFnP5JjOgJc)


All manufactures provides replacements for defective probes within the warranty period. But they do not provide replacement for physically damaged probes and no any service and repair for warranty expired probes. In most of time the user did not validate their probes on time and the defectiveness in probes remains unidentified. So it is necessary that all clinicians and hospital authority should validate their probe before warranty expires.

WHEN WE NEED TO REPAIR THE PROBE ???? Many of the probe damages are repairable at initial state. But in many cases the Practitio...